The facility stated 11 of the 25 staff members have been exposed to hydrogen peroxide from the v-pro 1 unit and experience irritation around the throat, eyes and nose.One employee sought and received medical treatment at the facility's emergency room after experiencing eye irritation.Following notification of the event, steris personnel from field service and engineering arrived at the facility and evaluated the unit.The evaluation concluded the v-pro 1 plus sterilizer is performing to specification.The unit successfully passed two pressure verification tests and a leak test.Additional test cycles were performed to simulate lumen and non lumen cycle runs.These test cycles successfully completed without alarms or cycle aborts.The room in which the v-pro 1 plus unit is located is not supplied with sufficient air exchanges per hour per aami st58:2013.On (b)(4) 2014, a conference call was held between steris personnel (product management, service engineering and r&d) and a representative from the user facility.During the call, the facility committed to addressing the aami standard for air exchanges per hour.
|