• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG & CO. KG CRANIOFIX 2 TITANIUM CLAMP 20MM; CRANIAL CLOSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG & CO. KG CRANIOFIX 2 TITANIUM CLAMP 20MM; CRANIAL CLOSURE Back to Search Results
Model Number FF492T
Device Problems Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Code Available (3191)
Event Date 04/08/2014
Event Type  malfunction  
Event Description
Country of complaint: (b)(6).Client complained: during surgery, the surgeon tried to tighten the cranial bone by 4 points, using ff492t.However, 2 of them could not be fixed properly and then the upper disks came off.
 
Manufacturer Narrative
Us reporting agent notified on: (b)(4) 2014.Mfg site evaluation: 2 of the 4 craniofixes have been cut too short.It can be clearly seen that the pin has been cut off too short where there are no more anti-clip serrations.The cause for the loosening of the plate is that the pin has been cut too short.This failure is user related.The mfg documentation has been checked and found to be according to the specification valid at the time of production.There are no indications for a material or mfg defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRANIOFIX 2 TITANIUM CLAMP 20MM
Type of Device
CRANIAL CLOSURE
Manufacturer (Section D)
AESCULAP AG & CO. KG
tuttlingen, de 7853 2
Manufacturer (Section G)
AESCULAP AG&CO KG
po box 40
tuttlingen 7850 1
GM   78501
Manufacturer Contact
michelle link
615 lambert pointe dr
hazelwood, MO 63042
3145515938
MDR Report Key4047637
MDR Text Key4809134
Report Number2916714-2014-00621
Device Sequence Number1
Product Code GXN
Combination Product (y/n)N
PMA/PMN Number
K972332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFF492T
Device Catalogue NumberFF492T
Device Lot Number51999407
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/23/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-