• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL S.A. OPTILENE 7/0 (0,5) 75CM 2XDR10FB CV RCP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL S.A. OPTILENE 7/0 (0,5) 75CM 2XDR10FB CV RCP Back to Search Results
Model Number C3095838
Device Problems Device Difficult to Setup or Prepare (1487); Device Operates Differently Than Expected (2913); Human-Device Interface Problem (2949)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2014
Event Type  malfunction  
Event Description
Country of complaint: (b)(6).Complaint states: there are two different problems at this product.One, the withdrawal of the needle is very heavy.You have the feeling that the package is chamber to the back.Due to this you couldn't grab the needle correctly.Two, the withdrawal of the needle is too strong.Due to this, the thread broke very easily during the withdrawal.
 
Manufacturer Narrative
Us reporting agent notified on: (b)(4) 2014.Manufacturing site evaluation: samples received: one unopened racepack.There are no previous complaints of this code batch.There are no units in oem stock.The withdrawal of the needle has been correct and the usual one.After holding the needle, the suture has been pulled out from the pack without difficulty and the usual one.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill oem requirements.Final conclusion: the complaint is not corresponding (not justified).Corrective/preventive actions: not applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTILENE 7/0 (0,5) 75CM 2XDR10FB CV RCP
Manufacturer (Section D)
B. BRAUN SURGICAL S.A.
rubi, barcelona
SP 
Manufacturer (Section G)
B. BRAUN SURGICAL S.A.
121 carretera de terrassa
rubi, barcelona
SP  
Manufacturer Contact
michelle link
615 lambert pointe drive
hazelwood, MO 63042
3145515938
MDR Report Key4047680
MDR Text Key4737772
Report Number2916714-2014-00633
Device Sequence Number1
Product Code FTL
Combination Product (y/n)N
PMA/PMN Number
K061704
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2019
Device Model NumberC3095838
Device Catalogue NumberC3095838
Device Lot Number114063V004
Was Device Available for Evaluation? No
Date Manufacturer Received07/10/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/05/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-