The customer reported the coulter lh 500 hematology analyzer generated ambient and lyse temperature errors.There was no exposure to open wounds or mucous membranes.No erroneous results were generated in connection with the reported event.There was no death, injury, change to patient treatment attributed to this event.
|
A beckman coulter (bec) field service engineer (fse) was dispatched to assess the instrument.The fse confirmed a leak from a hole in the black tubing through pinch valve (pv37).The leak consisted of approximately 20 mls of clenz that leaked from the analyzer onto the counter.The leak was not contained within the instrument.The fse replaced the tubing through pinch valve (pv37), resolving the leak.The leak had damaged the peltier circuit board and associated connector (p4).The fse replaced the peltier board and connector, resolving the ambient lyse and temperature errors.While onsite, the fse replaced the tubing on the center module, pump module, differential module, level sense module, and complete blood count (cbc) module as part of preventative measure.The fse verified the instrument and the instrument was returned to normal operation.The operator/fse was wearing the personal protective equipment (ppe) consisting of gloves, masks/protective eyewear and a lab coat at the time of the incident.(b)(4).
|