Reportable based on analysis completed on (b)(4) 2014.It was reported that the device failed to deploy.The 95% stenosed, 12mm in length, 8mm in diameter target lesion was located in the severely tortuous and severely calcified left internal carotid artery.The lesion had a bend over between 45 degree but less than 90 degree.A 190cm filterwire ez¿ was used for a carotid artery stenting procedure.During procedure, it was noted that the filter could not release normally.The physician removed the device and completed the procedure with another of the same device.There were no patient complications reported and the patient's status was stable.However, device analysis revealed that the filter bag was torn into 2 pieces.
|
Device evaluate by mfr.: investigation completed.The device was returned for evaluation.The device was returned without the packaging that contained the batch information.The protection wire and the delivery sheath were returned inside a plastic bag.The retrieval sheath was not returned.During visual and microscopic examination of the returned device, the radiopaque distal tip of the protection wire was found curved; it was not damaged, wavy, or stretched.The filter bag was found deployed, however, it was found torn into 2 pieces.Both pieces were still attached to the device and no missing or completely detached pieces were found.The tear was found at the tip of the filter bag and the material is still attached to the adhesive.Blood was found on the inside of the delivery sheath.Sheathing/unsheathing test was performed successfully.The filter bag was successfully retracted and then deployed.The peel away test was also performed successfully, no anomalies were observed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|