A product investigation was performed for the device.A simulated treatment was performed and completed without any failures or problems.The reported m01.17 alarm was not duplicated or verified.There was no evidence of fluid leaks found during visual inspection.All functional tests passed.A review of the device history records was performed and there were no unexpected variances, or adjustments, or comments noted in the device records.A search of the complaint tracking system showed one other complaint for m01.17 alarm during setup for this device, however, trending plan thresholds are not triggered.The clinical investigation indicates the pt's peritoneal dialysis nurse stated the peritonitis was due to touch contamination rather than the use of the cycler and there were no reported cassette fluid leaks.The peritonitis occurred on (b)(6) 2012 which was 15 days prior to the tech support call.The pt was treated with antibiotics for 8 days and then released from the hosp.The clinical investigation concludes peritonitis is typically due to touch contamination of the peritoneal dialysis catheter/tubing connection and not attributed to the cycler.This is a retrospective mdr being submitted by fresenius medical care as a remediation action committed to fda regarding all potential adverse events received between 03/21/2011 to 03/11/2013.
|