• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC., CRMD ZEPHYR XL SR; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC., CRMD ZEPHYR XL SR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5626
Device Problems Difficult to Interrogate (1331); Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/23/2014
Event Type  malfunction  
Event Description
It was reported that during follow-up, the pulse generator could not be interrogated.It was noted that there was trouble interrogating the device after implant as well.External monitoring showed pacing that suggested the device had entered into backup mode.On (b)(6) 2014, the device was successfully interrogated and backup operation was confirmed.A firmware download was performed and the device was restored to normal functionality.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR XL SR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jennie hamel
15900 valley view court
sylmar, CA 91342
8184933462
MDR Report Key4073808
MDR Text Key19720199
Report Number2017865-2014-16795
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2015
Device Model Number5626
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-