Model Number PM2112
Device Problem No Device Output (1435)
Patient Problems Cardiac Arrest (1762); Fainting (1847)
Event Date 07/08/2014
Event Type  Injury  
Event Description
It was reported that while receiving external electrical therapy for unrelated symptom, the patient fainted.The pulse generator exhibited loss of output.The patient was asystolic.The patient received external cardioversion.Following external therapies, normal device functions resumed.The device was explanted and replaced on (b)(6) 2014.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.Final analysis found normal pulse generator characteristics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jennie hamel
15900 valley view court
sylmar, CA 91342
8184933462
MDR Report Key4073809
MDR Text Key18539464
Report Number2017865-2014-16796
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2015
Device Model NumberPM2112
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/11/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age88 YR
Patient Weight75
-
-