(b)(4).Method: the complaint airvo humidifier was received at fisher & paykel healthcare (b)(6) and was visually inspected and performance tested at the customer settings for 24 hours with water bag attached to a water chamber for continuous water flow.The hospital had reported that the oxygen was disconnected from the oxygen port.We can conclude that the oxygen tube disconnected either from the oxygen supply or from the airvo barb connector.The hospital did not confirm what had caused the oxygen tubing to disconnect.If properly connected, the oxygen tubing fits firmly onto the barb connector and should not spontaneously disconnect.The 900pt402 oxygen inlet kit is sold as an accessory for the airvo 2 humidifier and allows the user to connect supplementary oxygen.Supplementary oxygen is connected to a barb connector, which is suitable for use with most common sizes of oxygen tubing.The airvo 2 user manual instructs the user to "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply." the user manual also states that the "airvo 2 is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases." and that "the unit is not intended for life support." result: visual inspection revealed no damage to the airvo or to the barb connector.During testing the unit performed with no errors or warnings displayed.Oxygen readings were observed to be normal at ambient.The power fail alarm was tested and found to be functioning.Conclusion: the subject airvo operated correctly during testing.We can conclude that the patient issue was caused by the disconnection of the oxygen supply.Our user instructions that accompany the airvo humidifier state that "the airvo 2 is for the treatment of spontaneously breathing patients (adult and pediatric > 3 kg) who would benefit from receiving high flow warmed and humidified respiratory gases." they also state: "the unit is not intended for life support.".
|