• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL; DISLODGER, STONE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL; DISLODGER, STONE, FLEXIBLE Back to Search Results
Model Number M0063903100
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/16/2014
Event Type  Injury  
Event Description
It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used during a transurethral lithotripsy procedure performed on (b)(6) 2014.According to the complainant, during the procedure, a stone cone was inserted through a 6fr ureteroscope and passed beside an impacted ureteral stone.After the device was deployed, approximately 16 centimeters of the distal portion of the device, including the coil, separated and migrated towards the kidney.The physician disintegrated the stone and retrieved the broken part of the device using a ureteroscopy grasper.The procedure was completed with another stone cone nitinol urological retrieval coil.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
The complainant was unable to report the upn and lot number; therefore, the manufacture date and expiration date are unknown.However, the complainant stated that the device was used prior to the expiration date.(b)(4).The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Type of Device
DISLODGER, STONE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
ACCELLENT ENDOSCOPY JUAREZ
31c butterfield trail
el paso TX 79906
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4082478
MDR Text Key4898924
Report Number3005099803-2014-03095
Device Sequence Number1
Product Code FGO
Combination Product (y/n)N
Reporter Country CodeIR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 08/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM0063903100
Device Catalogue Number390-310
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age40 YR
-
-