(b)(4).The complaint mr290hfv chambers have not been received at fisher & paykel healthcare for evaluation.Without the devices we are unable to determine what may have caused the issue as reported by the customer.However it is known that some cpap ventilators are capable of producing pressures in excess of 80cmh2o.The integrity of the chamber can be affected when pressures exceed 80cmh2o.Every mr290 chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.In addition, the pressure test is followed by a visual inspection of each chamber.Any chamber which fails either of these tests is rejected.The chambers would have met the required specification at the time of production.Our user instructions that accompany the mr290hfv state the following: - use of the mr290 above the maximum operating pressure [8kpa (~80cmh2o)] may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.
|