A steris service technician arrived at the facility, evaluated the v-pro max sterilizer and found the equipment operating to specification.No repairs were required; the reported event could not be duplicated.The steris technician confirmed that the employee was not wearing ppe, gloves, at the time of the event.The operator manual (6-11) states, "unload sterilization unit: steris recommends (in accordance with ansi/aami st58, 2005) wearing chemical -resistant gloves when using the sterilization unit.In addition, the operator manual (1-2) states, "danger - chemical injury hazard: any visible liquids in the chamber must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions." a steris account manager in-serviced the facility on the importance of wearing ppe and drying instruments completely before being placed in the v-pro max sterilizer.No further issues have been reported.
|