Model Number 106A3
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/20/2014
Event Type  Injury  
Event Description
Information received by medtronic indicated that the user was unable to begin a cryoablation procedure due to issues powering on the cryoconsole.Technical support was called during the case and were unable to resolve the issue with this cryoconsole.The case was aborted; the patient was under general anesthesia for the procedure and did not receive any therapeutic treatment.There were no patient complications.
 
Manufacturer Narrative
Investigation is in progress.Results of investigation will be submitted in a supplemental report.
 
Manufacturer Narrative
Technical support was notified of this occurrence.Upon review of complaint information, a service order was initiated to review the console performance on site.The reported issue was reproducible during site visit.The console's original power supply and 6a fuse were replaced and the issue was resolved; all other fuses were inspected and found to be intact.The console's mother board and watchdog board were also replaced.All catheter and console safety tests passed without anomaly and the console was ready for use.Suspected defective parts of the cryoconsole were returned for evaluation.The reported issue has been confirmed through testing and data analysis.The returned power supply failed the inspection due to power off.The returned mother board and watchdog board passed the performance test as per specification.This failure type shall be monitored and trended.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOCONSOLE
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland,qc H9H 5 H3
CA   H9H 5H3
Manufacturer Contact
voula radiotis
16771 chemin ste-marie
kirkland,qc H9H 5-H3
CA   H9H 5H3
5146941212
MDR Report Key4089775
MDR Text Key4739830
Report Number3002648230-2014-00154
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number106A3
Device Catalogue Number106A3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/26/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age00051 YR
Patient Weight75
-
-