No patient information provided as no patient was involved in this concern.Engineering analysis finds that in this particular case, it appears like user performed work list download to create patient.All the patient information will update correctly, however, may create a new record in the procedure step table and will create multiple records in the procedure table with the same study id.In this case, procedure step table contains more than one record with the same study id, application not loading the procedure table and then trying to load series table (child table) will result in the same error.(b)(6) 2014, a medtronic representative, following-up at the site, found no errors associated with dose reporting.Recommendation was made regarding reloading of the software.Software investigation was completed.This issue was documented in a medtronic software anomaly tracking database.
|