While on site, a beckman coulter (bec) field service engineer (fse) reported a leak from the customer's unicel dxh 800 coulter cellular analysis system.The volume of the leak was approximately 2 ml and came from a quick disconnect inside the vcsn module.The leak shorted out a cable connector and an electrical circuit board.The fse observed a burning smell, but did not observe any fire, sparks, or arcing.The fire department was not called and the laboratory was not evacuated.The instrument generated "solenoid is shorted" errors and became inoperable.The instrument operator was wearing a lab coat and gloves at the time of the leak.There were no reports of biohazard exposure to the leak.There were no erroneous results generated and patient treatment was not impacted in connection with this event.
|