• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. T-LINK DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. T-LINK DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 814851
Device Problems Charred (1086); No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/15/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, there was a blank screen on the t-link system.The power light emitting diode (led) was on.The device was not changed out, as the customer used paper charting for the case.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.Per the laboratory evaluation, the pc appears to power up normally, except the display (no backlight) never comes on.Discovered a wire that runs to the backlight inverter board was charred and broken.The product will be sent to service to be brought to manufacturers specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts nd/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-LINK DATA MANAGEMENT SYSTEM
Type of Device
T-LINK DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4105543
MDR Text Key16303465
Report Number1828100-2014-00782
Device Sequence Number1
Product Code DXJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number814851
Device Catalogue Number814851
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/05/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-