It had been reported that revision surgery was performed due to a broken expedium rod and possible infection.During the procedure, the surgeon had difficulty removing expedium single inner set screws due to stripping of the inner hex features and decided not to remove them.The tips of two moss miami si final tighteners were stripped during the procedure.Medwatch reports were filed for the implant devices due to the possible infection (including the broken rod and six expedium single inner set screws), and two moss miami si final tighteners that were stripped.In addition, external damage to the expedium si intermediate tightener, (b)(4), wasn¿t apparent when the complaint was reported or upon receipt of the instrument for evaluation.Further investigation of the returned intermediate tightener determined that the hex lobes on the distal end of the instrument were torn.As a result, this mfg.Medwatch report 1526439-2014-11913 is being submitted for the expedium si intermediate tightener.
|
The expedium si intermediate tightener was returned for evaluation.Visual inspection found some of the lobes partially torn off.This damage is sufficient enough to interfere with the functionality of the instrument.A review of the device history record (dhr) identified no issues during the manufacturing and release of this product that could have contributed to the problem reported by the customer.The product was released accomplishing all quality requirements.A review of the complaint trend analysis for this device code found no related complaints.The root cause cannot be positively determined based on the samples and information provided.However a potential root cause may be wear and tear after repeated use as the driver are greater than 2 years old.No corrective or preventative action (capa) is necessary at this time as there has been no observed systemic trend.Therefore, the complaint will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|