• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVATORS MEDIVATORS DSD-201; AUTOMATED ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVATORS MEDIVATORS DSD-201; AUTOMATED ENDOSCOPE REPROCESSOR Back to Search Results
Model Number DSD-201
Device Problems Misconnection (1399); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/28/2014
Event Type  No Answer Provided  
Event Description
The case states that facility is using hookups with their endoscopes that have been completely altered/modified by user.The hookups are not validated for use in any medivators automated endoscope reprocessors.Potential for cross contamination.
 
Manufacturer Narrative
The case states that facility is using hookups/aer endoscope connectors that have been completely altered/modified by user.The hookups are not validated for use in any medivators automated endoscope reprocessors.Potential for cross contamination.This was reported by medivators field service engineer via field observation report and photographs.To date, there have been no reported injury or illness.This complaint will be maintained within medivators complaint system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDIVATORS DSD-201
Type of Device
AUTOMATED ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
MEDIVATORS
14605 28th ave n
minneapolis MN 55447
Manufacturer (Section G)
MEDIVATORS
14605 28th ave n
minneapolis MN 55447
Manufacturer Contact
megan dickey
14605 28th ave n
minneapolis, MN 55447
7635533327
MDR Report Key4112940
MDR Text Key4918453
Report Number2150060-2014-00037
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K914145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Notification
Type of Report Initial
Report Date 09/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberDSD-201
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/28/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-