• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY TULSA DENTAL SPECIALTIES E3 TORQUE CONTROL MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY TULSA DENTAL SPECIALTIES E3 TORQUE CONTROL MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number E3KIT
Device Problems Overheating of Device (1437); Retraction Problem (1536)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/26/2014
Event Type  malfunction  
Event Description
In this event it was reported that an e3 motor failed to auto-reverse and was also overheating; there is no indication that injury resulted.
 
Manufacturer Narrative
Because failure of the auto-reserve function could possibly lead to file separation, which could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the issue would be likely to cause/contribute to a serious injury should it recur.A such, this event meets criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E3 TORQUE CONTROL MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY TULSA DENTAL SPECIALTIES
johnson city TN
Manufacturer Contact
helen lewis
221 w. philadelphia st., ste 60
susquehanna commerce center w.
york, PA 17401
7178457511
MDR Report Key4117946
MDR Text Key4702252
Report Number9611053-2014-00003
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103653
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 06/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberE3KIT
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/21/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-