• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS PERFORATOR DRIVER WITH HUDSON END; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS PERFORATOR DRIVER WITH HUDSON END; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number CSR60
Device Problems Positioning Problem (3009); Temperature Problem (3022)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
Customer reported that the device heats up and it is not spinning when the drill is running.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Additional narrative: synthes is submitting this report as a result of remediation activities related to synthes service and repair files legacy review/remediation protocol-complaint handling and mdr reporting.Device listed in this report is used for treatment, not diagnosis.Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr.During pre-repair assessment performed by technician, it was noted that the gears were damaged due to normal wear.The device was repaired by replacing several components.Mfr date: 12/2010.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORATOR DRIVER WITH HUDSON END
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4126354
MDR Text Key4949708
Report Number1045834-2014-13074
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK965080
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/04/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCSR60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2012
Is the Reporter a Health Professional? No
Date Manufacturer Received12/04/2012
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-