Catalog Number 6605632
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/02/2014
Event Type  malfunction  
Event Description
Customer reported conductivity was drifting high daily when using the coulter lh 750 hematology analyzer.A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the analyzer.There was no report of erroneous test results associated with this event.There was no death, injury or affect to user or patient treatment.
 
Manufacturer Narrative
On (b)(4) 2014, a beckman coulter field service engineer (fse) evaluated the analyzer and confirmed that the conductivity was drifting daily due to a defective rf preamp detector card.The fse replaced the rf preamp detector card resolving the issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s kraemer blvd
brea, CA 92821
7149614941
MDR Report Key4126719
MDR Text Key17995757
Report Number1061932-2014-02460
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-