• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 DOCKING STATION (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 DOCKING STATION (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Catalog Number 0702014000
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 09/03/2014
Event Type  malfunction  
Event Description
It was reported that while docking a rover at the user facility, the water supply hose leaked onto the power module, causing the docker to produce smoke.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
The reported smoking of the device was confirmed by a manufacturer field service technician.Upon inspection, it was found that the fresh water hose was leaking onto the power plug module, which can cause the reported event.The device was repaired at the user facility by the manufacturer field service technician and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPTUNE 2 DOCKING STATION (120V)
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key4127565
MDR Text Key4795441
Report Number0001811755-2014-03423
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132671
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0702014000
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/03/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-