STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE 2 DOCKING STATION (120V); APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Catalog Number 0702014000 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/03/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that while docking a rover at the user facility, the water supply hose leaked onto the power module, causing the docker to produce smoke.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The reported smoking of the device was confirmed by a manufacturer field service technician.Upon inspection, it was found that the fresh water hose was leaking onto the power plug module, which can cause the reported event.The device was repaired at the user facility by the manufacturer field service technician and returned to service.
|
|