Lot, consisted of 10 pieces, was received from supplier and purchased by customer on 11jun2014.Visual examination found both cutting tubes contained residue in the interior, which promoted the corrosion of the device.Specifically, corrosion was found in the transition area from the pipe to the driver and the area of the coded pins.Supplier has eliminated the groove in the transition area in order to greatly reduce residue accumulation.In addition, supplier has improved the cutting tubes near the insertion and tightened tolerances.Change implemented under supplier capa (b)(4).Since the market launch on 13jun2008 no similar complaints of cutting tube corrosion have been received.Instructions for use (ifu), ga-a245, contained sufficient instructions.In the risk assessment, e5-1, potential risks of the cutting tube were taken into account and had an acceptable risk rating.Richard wolf (b)(4) considers this matter closed.However, in the event (b)(4) receives any additional information a follow up report will be submitted to fda.Richard wolf medical instruments corporation (rwmic) submitting report on behalf of (b)(4).
|