Discordant, falsely low calcium results were obtained on patient samples on an advia 1800 instrument.The discordant results for two of the patient samples were reported to the physician(s), but were corrected immediately.The samples were repeated on the same instrument and an alternate instrument and all resulted higher on both.There are no reports of patient intervention or adverse health consequences due to the discordant calcium results.
|
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse cleaned and flushed the wash up down and dilution wash up down, replaced a filter and seals on the sample probe and dilution probe pumps, and cleaned and calibrated all probes and mixers.The cse ran a diagnostics test, which passed.The cause of the discordant calcium results is unknown.The instrument is performing according to specifications.No further evaluation of this device is required.
|