A discordant, falsely elevated troponin result was obtained on one patient sample on an advia centaur cp instrument.The discordant result was not reported to the physician(s).The sample was repeated on the same instrument and another advia centaur cp instrument, resulting lower.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated troponin result.
|
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse replaced ring bearings, waste check valve , wash station manifold, luminometer, rotary pumps, sample syringe and diluter, reagent syringe and diluter and tubing.The cse also replaced the ionizer fan replaced to resolve signal errors.Patient samples were run, resulting as expected.The cause of the discordant, falsely elevated troponin result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|