• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE; APPARATUS, EXHAUST, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE; APPARATUS, EXHAUST, SURGICAL Back to Search Results
Catalog Number 0700001000
Device Problem Sparking (2595)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2014
Event Type  malfunction  
Event Description
It was reported that during a surgical procedure at the user facility the device was sparking.The procedure was completed successfully.No delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
The reported sparking was confirmed by a manufacturer field repair technician.Upon disassembly, it was found that the power plug contained loose wires, which can lead to the reported event.The device was repaired at the user facility and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE
Type of Device
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key4134789
MDR Text Key4795579
Report Number0001811755-2014-03445
Device Sequence Number1
Product Code FYD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0700001000
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/05/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-