• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI TENDRIL STS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI TENDRIL STS Back to Search Results
Model Number 2088TC/58
Device Problems Under-Sensing (1661); Difficult to Advance (2920)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/28/2012
Event Type  Injury  
Event Description
It was reported that a lead was undersensing.The lead was repositioned, but the stylet would not fully insert into the lumen of the lead.The lead was ultimately replaced.
 
Manufacturer Narrative
This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.Final analysis showed that the inside inner coil was clogged with blood, which was causing stylet insertion difficulty.Electrical tests of the lead found no anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Manufacturer (Section D)
ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI
15900 valley view ct.
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISI
15900 valley view ct.
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view ct.
sylmar, CA 91342
8184932621
MDR Report Key4143680
MDR Text Key4913894
Report Number2017865-2014-05754
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Model Number2088TC/58
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/08/2013
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date12/28/2012
Device Age6 MO
Event Location Hospital
Date Manufacturer Received12/28/2012
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
-
-