• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC, CRMD TENDRIL SDX; PERMANENT PACEMAKER ELECTRODE, DTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC, CRMD TENDRIL SDX; PERMANENT PACEMAKER ELECTRODE, DTS Back to Search Results
Model Number 1688TC/52
Device Problems Decreased Sensitivity (2534); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2012
Event Type  Injury  
Event Description
It was reported that during a routine follow up, the right ventricular lead exhibited increased capture threshold and decreased sensing.A chest x-ray revealed that the lead had dislodged.The lead was repositioned and remained implanted.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.This historical complaint is being filed as part of a retrospective review of complaint files in response to a recent fda inspection.There is no change to the actual performance of the product and this report only represents an enhancement to the reporting criteria going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL SDX
Type of Device
PERMANENT PACEMAKER ELECTRODE, DTS
Manufacturer (Section D)
ST. JUDE MEDICAL, INC, CRMD
15900 valley view ct.
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC
15900 valley view ct.
sylmar CA 91342
Manufacturer Contact
cary lawler
15900 valley view ct.
sylmar, CA 91342
8184932621
MDR Report Key4144361
MDR Text Key4767755
Report Number2017865-2014-06790
Device Sequence Number1
Product Code DTS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2014
Device Model Number1688TC/52
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date11/01/2012
Device Age15 MO
Event Location Hospital
Date Manufacturer Received11/01/2012
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
-
-