• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARJOHUNTLEIGH, INC. ROTORESTDELTA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJOHUNTLEIGH, INC. ROTORESTDELTA Back to Search Results
Device Problems Misconnection (1399); Device Operational Issue (2914)
Patient Problem Death (1802)
Event Date 09/05/2014
Event Type  Death  
Event Description
It has been reported to arjohuntleigh, that a resident nurse discovered an exposed bracket under the patient's sacrum area, when she went to chang and clean under the patient.A service order was issued to collect asset serial number (b)(4), and deliver a replacement bed.However, the patient died of apparently unrelated causes while the replacement bed was in route.
 
Manufacturer Narrative
This report is being filed by the manufacturer arjohuntleigh, inc.Please not that previous medwatch reports for this product may have been submitted from the manufacturing site kinetic concepts, inc.As of (b)(4) 2012, complaints related to this product are to be handled by arjohuntleigh, inc.When reviewing similar reportable events for rotoprone rotorestmi and reotorest delta therapy system, we have not been able to found any similar fault description compared to the situation investigated here: patient lying on the device with reversely assembled hatches.There is no trend observed for reportable complaints with this failure for rotorest delta therapy system.Based on the collected information, it has been determined that the contributory factor in the patient's death was advanced patient medical condition not a device condition.Please note that the patient's comorbidities included the following: acute respiratory distress syndrome (ards), septic shock and multiple organ dysfunction syndrome.The patient had some minor indentation on the skin across the buttocks area that did not require any medical intervention.The patient did not sustain any serious injuries from the hatches being misplaced.The patient's death was unrelated to the bed or to the misplaced hatches.Unfortunately, we have been unable to clarify where or when the hatch assembly was reversed.Up to this point ther is no indication that the device caused or contributed to the patient death.In summary, the device was being used, when the event occurred, for the patient treatment, the device contributed to the event since the patient was lying on the device, however, we could not foind any indication that the device actively caused or contributed to the event, it was merely present.Nevertheless as the hatches were reversed on the bed, it failed to meet its specification.Unfortunately, the patient who was lying on the bed, died while the replacement bed was in route.Given the circumstances and no number of similar events, we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTORESTDELTA
Manufacturer (Section D)
ARJOHUNTLEIGH, INC.
12625 wetmore road
suite 308
san antonio TX 78747
Manufacturer (Section G)
ARJOHUNTLEIGH, INC
12625 wetmore road
ste 308
san antonio TX 78747
Manufacturer Contact
pamela wright
12625 wetmore
ste 308
san antonio, TX 78747
2102787040
MDR Report Key4151459
MDR Text Key4914074
Report Number3010048749-2014-00081
Device Sequence Number1
Product Code IKZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial
Report Date 09/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/05/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age78 YR
Patient Weight102
-
-