Device 1 of 5.Reference mfr.Report#: 1627487-2014-23652, reference mfr.Report#: 1627487-2014-23654, reference mfr.Report#: 1627487-2014-23658, reference mfr.Report#: 1627487-2014-23659.It was reported the patient is experiencing an uncomfortable surge from the ipg to his legs.The sjm representative met with the patient and determined the patient's ipg is inoperable.In addition, no communication could be established between the patient's ipg and external devices.The patient's programmer also reflects a communication error.The patient attempted to recharge the ipg; however, he was unsuccessful.The patient is without stimulation.Surgical intervention will be undertaken as the next course of action.
|