• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS CANADA CORPORATION 1227 CART WASHER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CANADA CORPORATION 1227 CART WASHER Back to Search Results
Device Problems Fumes or Vapors (2529); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/09/2014
Event Type  No Answer Provided  
Event Description
The user facility reported that a cycle aborted with a message of "alarm #44-unload door open during cycle".The employee opened the unload door and vapor released subsequently setting off the sprinkler system.No fire alarms were triggered and the fire department was not dispatched.No evacuations were reported.No injuries or procedural delays/cancellations were reported.
 
Manufacturer Narrative
A steris service technician arrived onsite and found the unload door to be open and the cart washer to be turned off via the emergency stop switch on the unload side of the washer.The technician ran several cycles and could not duplicate the reported event.The technician replaced the unload door switch, ran test cycles and confirmed the washer to be operational.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1227 CART WASHER
Type of Device
WASHER
Manufacturer (Section D)
STERIS CANADA CORPORATION
490 armand-paris
quebec, quebec GIC 8A3
CA  GIC 8A3
Manufacturer (Section G)
STERIS CANADA CORPORATION
490, armand-paris
quebec, quebec GIC 8A3
CA   GIC 8A3
Manufacturer Contact
kathryn cadorette
5960 heisley road
mentor, OH 44060
4403927231
MDR Report Key4152736
MDR Text Key4787618
Report Number9680353-2014-00083
Device Sequence Number1
Product Code FLH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-