Device Problems
Fumes or Vapors (2529); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/09/2014 |
Event Type
No Answer Provided
|
Event Description
|
The user facility reported that a cycle aborted with a message of "alarm #44-unload door open during cycle".The employee opened the unload door and vapor released subsequently setting off the sprinkler system.No fire alarms were triggered and the fire department was not dispatched.No evacuations were reported.No injuries or procedural delays/cancellations were reported.
|
|
Manufacturer Narrative
|
A steris service technician arrived onsite and found the unload door to be open and the cart washer to be turned off via the emergency stop switch on the unload side of the washer.The technician ran several cycles and could not duplicate the reported event.The technician replaced the unload door switch, ran test cycles and confirmed the washer to be operational.No additional issues have been reported.
|
|