(b)(4).Lake region medical reviewed the device history records relative to the manufacturing, inspecting and packaging of lot 10324295.The device history record review also included a review of the certificate of conformance received from specialty coatings systems and nitinol devices and components, along with lake region medical¿s internal receiving inspection records for the stents issued to the complaint lot.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.Based on the information, the event could not be confirmed.The product was not returned for analysis; however, a review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Therefore, no corrective actions will be taken at this time.(b)(4).
|
Additional information: the microcatheter was exchanged over a guidewire, retaining target position.There was no clinically significant delay in the procedure due to the issue.An adequate continuous flush was maintained through the catheter.There was no difficulty during introduction of the catheter over the guidewire prior to the attempted use of the device.There was no difficulty tracking the microcatheter to the target site prior to introduction of the stent.There were no kinks on the microcatheter before or after the difficulty.After further information was received, the event does not meet criteria for medwatch reporting as there was no loss in target vessel position.Therefore, please note that no further information will be forthcoming for this report.
|