• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM Back to Search Results
Catalog Number 425-00
Device Problem No Display/Image (1183)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 09/18/2014
Event Type  malfunction  
Manufacturer Narrative
A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.Per dhr (device history record) the product concha neptune, serial #(b)(4) was manufactured on 11/03/2008.The dhr investigation did not show issues related to complaint.A document assessment (fmea) was conducted and no changes were required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the defect.At this time since the device sample is not available, it is not possible to determine the source of the defect reported.The complaint can not be confirmed since the sample was not available for investigation, therefore, it is not possible to determine the root cause for the defect reported and a corrective action for it.If the device sample becomes available at a later date, this complaint will be updated accordingly.Teleflex will continue to monitor and trend relating complaints.
 
Event Description
The event is reported as: the customer alleges that the temperature readings will not display after the blue heated wire pigtail is connected.The issue was detected prior to patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HEATED HUMIDIFICATION SYSTEM
Manufacturer (Section D)
TELEFLEX
rtp NC 27709
Manufacturer (Section G)
PROLONGACION MISION EUSEBIO
kino # 1316
rancho el descanso
Manufacturer Contact
margie burton, rn
p.o. box 12600
durham, NC 27709
9194334965
MDR Report Key4156066
MDR Text Key18711885
Report Number3003898360-2014-00802
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 09/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number425-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/18/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-