SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW STRATUS CS Back to Search Results
Catalog Number 10444834
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2014
Event Type  malfunction  
Event Description
Customer reported discordant troponin results on the instrument.There was no report of injury due to this event.
 
Manufacturer Narrative
Customer indicated that they will not be able to send back a sample since the sample was used up rerunning the test.The cause for the discordant troponin results is unknown.
 
Manufacturer Narrative
The customer described a situation consistent with heterophilic interference with the stratus cs method.The sample was not able to be returned for evaluation so the presence of a non-specific binding interference such as a heterophilic antibody cannot be verified.Based on the review of message log there was no instrument or reagent issues could be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATUS CS
Type of Device
STRATUS CS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
glasgow business community
newark DE 19702
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
glasgow business community
newark DE 19702
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4156750
MDR Text Key4933996
Report Number1217157-2014-00153
Device Sequence Number1
Product Code DAP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051650
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10444834
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-