Model Number 2067
Device Problems Electrical /Electronic Property Problem (1198); Communication or Transmission Problem (2896); Delamination (2904)
Patient Problem No Patient Involvement (2645)
Event Date 08/05/2014
Event Type  malfunction  
Event Description
It was reported that the radiofrequency programmer head originally returned as it was no longer needed subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
 
Manufacturer Narrative
This report is based solely on device analysis.No information to suggest a device related adverse event or product problem was received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: during analysis the head failed all uplink amplitude tests and it was further noted that the cable was twisted and that the label was delaminating.The head was being sent on for repair.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4158350
MDR Text Key21173327
Report Number2182208-2014-02521
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2067
Device Catalogue Number2067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received08/27/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-