Model Number 5076-45
Device Problem Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/02/2014
Event Type  malfunction  
Event Description
It was reported that during the implant procedure, the stylet was not advancing smoothly, which made the lead manipulation difficult.The lead was positioned in the atrium with some effort but it was eventually discontinued due to its poor pacing and sensing thresholds.The lead was attempted and not used.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary: the full lead was returned, analyzed and the outer insulation of the lead was extrinsically breached due to a cut.The outer insulation of the lead was observed to have blood ingression.Visual summary analysis of the lead indicated damage at implant.Stylet insertion test result was passed.Visual analysis found outer insulation breached/ extrinsic and blood ingression.The breach insulation did contribute to the sensing complaint.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS
Type of Device
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
nasaco tech centre, 49 changi
singapore
SG 
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key4161427
MDR Text Key16302996
Report Number3008973940-2014-00374
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/21/2015
Device Model Number5076-45
Device Catalogue Number5076-45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-