The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the full lead was returned, analyzed and the distal and proximal conductors of the lead were obstructed due to a stylet/guidewire stuck in the lumen.Visual summary analysis of the lead indicated damage at implant.The lead was returned with the competitor guidewire stuck in the lead.It is likely that the green coating on the competitor guidewire became ensnared within the distal and proximal conductors causing the guidewire to become stuck.When trying to remove the guidewire, the lead was damaged during the attempted implant.(b)(4).
|