Model Number ADVIA CENTAUR XP
Device Problems False Device Output (1226); High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/15/2014
Event Type  malfunction  
Event Description
A discordant, falsely elevated troponin result was obtained on one patient sample on an advia centaur xp instrument.It is unknown if the discordant result was reported to the physician(s).The sample was initially flagged by the instrument and did not generate a value.The sample was then repeated on the same instrument, resulting falsely high.The sample was repeated on an alternate advia centaur instrument, resulting lower.It is unknown if the corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated troponin result.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse cleaned the luminometer and emptied and cleaned the acid and base reservoirs.Quality controls were run, resulting within range.The cause of the discordant, falsely elevated troponin result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS
manufacturing limited
registration number: 8020888
chapel lane, swords, co., dublin
EI  
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown 10591
9145242687
MDR Report Key4163601
MDR Text Key4872507
Report Number2432235-2014-00596
Device Sequence Number1
Product Code MMI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR XP
Device Catalogue Number078-A011-03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/15/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/28/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-