MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS SUPER SUCKER SHORT, X50 BULK PACKED; INSTRUMENT, SURGICAL, DISPOSABLE - SUCTION TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS SUPER SUCKER SHORT, X50 BULK PACKED; INSTRUMENT, SURGICAL, DISPOSABLE - SUCTION TIP Back to Search Results
Catalog Number 63-SS2-1
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2014
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.A visual assessment was performed which confirmed that the filter locking tab was missing.The assignable root cause was determined to be due to the tab getting snagged and pulled off during processing.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported from canada that during a total hip surgery, it was observed that the tab was broken on the suction nozzle of the super sucker device.There were no delays to the surgical procedure.It was not reported if a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The reported stated that the event occurred during the month of (b)(6) 2014; however, the exact date of the event was unknown.The reporter stated that the device was used in a canadian facility.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER SUCKER SHORT, X50 BULK PACKED
Type of Device
INSTRUMENT, SURGICAL, DISPOSABLE - SUCTION TIP
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key4163659
MDR Text Key13341575
Report Number1045834-2014-13133
Device Sequence Number1
Product Code KDC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK822065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number63-SS2-1
Device Lot NumberG483087436
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received05/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-