Model Number 20
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 09/16/2014
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that their device would not detect a quik-combo therapy cable when connected properly and attached to a test load.The customer further advised that this occurred with two different quik-combo therapy cables and would lose connection when the cable was wiggled near the therapy connector assembly.As a result, a patient may not be able to be detected if defibrillation therapy was needed.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4).Physio-control examined the customer's device but was unable to duplicate or verify the reported issue.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.The cause of the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4167695
MDR Text Key5094850
Report Number3015876-2014-01208
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2014
Is the Reporter a Health Professional? Yes
Device Age6 YR
Event Location Hospital
Date Manufacturer Received09/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-