It was reported that this event involved a pt with chronic renal failure.The catheter was inserted on (b)(6) 2014 into the pt's right femoral vein for urgent dialysis due to an inadequate shunt.After a few successful dialyses ((b)(6) 2014), it was observed that the catheter was kinked 19cm from the catheter tip.On subsequent dialysis, blood removal was impossible and there was also a slight resistance during blood transmission.The physician checked the catheter condition by ct image and confirmed vein perforation by the catheter tip around the pt's hip.The catheter was immediately removed and astriction (compression to arrest hemorrhage) was applied.A new catheter was inserted into the right internal jugular vein and treatment was successfully continued.The removed catheter was discarded in the ctr.It was noted that an x-ray was taken during insertion and no problems were observed.The pt had tortuous vessels.There were body movements in the pt.
|