• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL EXPRESS 4; STATSPIN CENTRIFUGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL EXPRESS 4; STATSPIN CENTRIFUGE Back to Search Results
Catalog Number X00-005331-001
Device Problems Fire (1245); Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 08/21/2014
Event Type  malfunction  
Event Description
The customer's biomed reported observing a spark, smoke, and a quickly self-extinguished "small electrical fire" upon power-up after replacing two internal components in accordance with detailed repair instructions.
 
Manufacturer Narrative
Per the biomed: he replaced the board and power harness, per the provided written and video instructions.Upon power up, the board sparked and had a "small electrical fire" and smoke was observed.He quickly disconnected the power.The flame was self-extinguished.He was wearing personal protective equipment (goggles, lab coat, and gloves).There was no injury (and no medical attention was sought) or damage to property; and we have requested the unit back for failure investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS 4
Type of Device
STATSPIN CENTRIFUGE
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton ave.
chatsworth, CA 91311
8185277272
MDR Report Key4172638
MDR Text Key5095131
Report Number2023446-2014-00150
Device Sequence Number1
Product Code JQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/21/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberX00-005331-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/21/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-