Catalog Number 031-33J
Device Problems Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2014
Event Type  malfunction  
Event Description
The event is reported as: the distributor (b)(4) alleges that a pin hole was observed on the package during incoming inspection.A total of (15) packages were observed with this reported issue.
 
Manufacturer Narrative
No visual or functional inspection can be performed since the device sample is not available for evaluation.The device history record (dhr) for lot#02b1400691 was reviewed was found that an ncmr was opened for damage during transit.Complete lot was inspected 100% visual inspection.The product was re-inspected and re-packed, however this type of defect is easy to detect during the manufacturing of the product since is verified at 100%.Additionally, per internal procedure, distribution center performs 100% inspection as well.Dhr shows that the product was packed and inspected according to our specifications.During investigation it was found that component tfx-000354 (sheet foam) batch (b)(4) was used during the manufacturing process, this sheet foam was implemented to prevent product damage.No conclusion can be established at this time based on the lack of device sample.It is necessary to evaluate the physical sample in order to perform a proper investigation.If the product sample becomes available this complaint will be re-opened.However, manufacturing and r&d departments are evaluating the current packaging design, a proposal of a new packaging method has been identified, and therefore a process/design has been initiated.Personnel involved in the manufacturing process have been notified on this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033, STERILE, JAPANESE
Type of Device
NEBULIZER ADAPTOR
Manufacturer (Section D)
TELEFLEX
rtp NC 27709
Manufacturer Contact
margie burton, rn
po box 12600
durham, NC 27709
9194334965
MDR Report Key4172813
MDR Text Key5157706
Report Number3004365956-2014-00368
Device Sequence Number1
Product Code CCQ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 09/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number031-33J
Device Lot Number02B1400691
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-