• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/08/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4) for the evaluation result of wire broke.Investigation results analysis of the returned rx cytology brush did not confirm the reported complaint.The brush, working length, and the t-fitting was taken off and not returned.The distal end of the handle cannula was found to be broken.Additionally, the wire was found to be bent.Therefore, the most probable root cause is operational context since the device was difficult to use and was forcefully used and made the wire bend and break.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.A search of the complaint database confirmed that no similar complaints exist for the specified batch.
 
Event Description
Note: this event, as reported, did not reflect an mdr reportable scenario; however, evaluation of the returned device revealed an mdr-reportable malfunction.This manufacturer report is being submitted based on the evaluation results.It was reported to boston scientific corporation that an rx cytology brush was used in the pancreatic duct during a cytology procedure performed on (b)(6) 2014.According to the complainant, during the procedure, when the device was actuated, the handle cannula bent.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4174118
MDR Text Key21221025
Report Number3005099803-2014-03329
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/01/2016
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number16941743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/20/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-