• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; CIRCULATORY ASSIST DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; CIRCULATORY ASSIST DEVICE Back to Search Results
Catalog Number 595000-001
Device Problem Improper Alarm (2951)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2014
Event Type  malfunction  
Event Description
The customer reported that the freedom driver exhibited an irreversible fault alarm while supporting a pt.The customer also reported that the pt said he was asymptomatic.The pt was subsequently switched to the backup freedom driver without adverse impact.Although the freedom driver exhibited fault alarm, the driver continued to function, as intended.
 
Manufacturer Narrative
This alleged failure mode poses a low risk to the pt because it does not prevent the freedom driver from performing its life-sustaining functions.The freedom driver will be returned to syncardia for eval.The results of the investigation will be provided in a supplemental mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
CIRCULATORY ASSIST DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer Contact
carole marcot, vp
1992 e. silverlake rd.
tucson, AZ 85713
5205451234
MDR Report Key4178285
MDR Text Key5155176
Report Number3003761017-2014-00146
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number595000-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age36 YR
-
-