BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605632
Device Problems Disconnection (1171); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2014
Event Type  malfunction  
Event Description
The customer reported a leak from a coulter lh 750 hematology analyzer.The volume of the leak was one cup and was not contained within the instrument.The instrument operator was wearing gloves, a face shield, and a lab coat at the time of the leak.There were no reports of direct contact with the leak.There were no erroneous results generated and patient treatment was not impacted in connection with this event.The customer found the self-cleaning line on the differential shear valve vcl85/126 was disconnected.The tubing was trimmed and reattached.The instrument was able to run without any leaks.
 
Manufacturer Narrative
The customer was able to resolve the leak.Beckman coulter (bec) field service was not dispatched to the site.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key4182645
MDR Text Key5161540
Report Number1061932-2014-02540
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Other Device ID NumberSW VERSION 2D3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/18/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-