Device evaluated by mfr: device was returned for analysis.The device is severely damaged at the distal section.The distal section is kinked at 1.7cm from the tip and is twist at 9cm from the tip, and the ring 2 is deformed and the adhesive is broken, there is blood residues inside the device, the center support is protruding out, ripping the distal section at 1.5cm from the tip.Also, there are chars in the rings 1 and 2.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|