• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 09/03/2014
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that a service indicator was displayed on their device and an event code was logged.There was no patient use associated with the reported event.Upon evaluation of the device, physio-control observed that the device had a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.
 
Manufacturer Narrative
(b)(4): physio-control evaluated the customer¿s device and verified the reported failure.Physio-control replaced the therapy pcb assembly.Thereafter proper device operation was observed through functional and performance testing and the device was returned to the customer for use.Physio-control further evaluated the removed therapy pcb assembly.It was determined that the cause of the reported issue was due to a failure of filter, designator fl8 on the therapy pcb assembly.The filter was observed to be physically broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4185111
MDR Text Key5020119
Report Number3015876-2014-01234
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2014
Is the Reporter a Health Professional? Yes
Device Age7 YR
Event Location Hospital
Date Manufacturer Received09/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/08/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-