MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number DIMENSION VISTA 1500
Device Problems False Device Output (1226); Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/29/2014
Event Type  malfunction  
Event Description
A discordant, falsely low vancomycin result was obtained on one patient sample on a dimension vista 1500 instrument.The discordant result was reported to the physician(s), who questioned it.The sample was repeated on the same instrument and on another dimension vista instrument, resulting higher.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low vancomycin result.
 
Manufacturer Narrative
The customer contacted the siemens customer care center (ccc).Ccc evaluated the instrument data and determined that quality controls were within range at the time of event.The cause of the discordant, falsely low vancomycin result is unknown, as the sample resulted as expected upon repeat testing on the same instrument.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 1500
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
brookfield CT 06804
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
brookfield CT 06804
Manufacturer Contact
cassandra kocsis
511 benedict ave
tarrytown, NY 10591
9145242687
MDR Report Key4186180
MDR Text Key18796151
Report Number1226181-2014-00519
Device Sequence Number1
Product Code JJE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K051087
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIMENSION VISTA 1500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/29/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/21/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-