• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2014
Event Type  malfunction  
Event Description
The customer reported the electrode in the white blood cell (wbc) bath was defective on the unicel dxh 800 coulter cellular analysis system in their laboratory.A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the analyzer.There were no erroneous test results with this event.There was no death, injury or affect to user or patient treatment.
 
Manufacturer Narrative
(b)(4) 2014, a beckman coulter field service engineer (fse) evaluated the system and found the wbc (white blood cell) bath electrode was defective.The fse replaced the wbc bath (which houses the electrode) to repair the instrument.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s kraemer blvd
brea, CA 92821
7149614941
MDR Report Key4186955
MDR Text Key5096484
Report Number1061932-2014-02656
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081930
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number629029
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-